The lot number is always a six-digit number, where the first digits indicate the manufacturing sub-batch during the month and the last three digits indicate the month and year of the manufacture (MMY). The last three digits indicating month and year of manufacture are most significant. For example, a lot number ending in 086 indicates that this lot was manufactured in August 2006. -All lots manufactured in July, 2006 (076) or earlier are affected. -Compact Absorbers from lot numbers 133086 and 136086, manufactured in August, 2006, should be isolated and destroyed. All other lots manufactured in August 2006 (086) are acceptable for use. -All lots manufactured in September 2006 (096) or later are acceptable for use.
Worldwide, including USA, Austria, Australia, Belgium, Canada, Croatia, Germany, Denmark, Finland, France, Greece, Iceland, Ireland, Italy, Japan, South Korea, Kuwait, Netherlands, Norway, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom.
Descripción del producto
Datex Ohmeda Compact Absorber, Disposable, REF 427002100, white to violet and REF 427002000, Pink to white. Used with GE Healthcare ADU anesthesia systems. GE Healthcare Finland Oy, Helsinki, Finland +358 10 39411.