PROSTIVA RF

Fabricante

  • Empresa matriz del fabricante (2017)
  • Comentario del fabricante
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    SATP

5 dispositivos con un nombre similar

Más información acerca de la data acá

  • Modelo / Serial
  • Descripción del producto
    Medical electronics / Electromedical devices
  • Manufacturer
  • Modelo / Serial
    All lot numbers are included.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution -- AL, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, PR, RI, SC, SD, IN, TX, UT, VA, VT, WA, WI, WY, and WV and countries of Australia, New Zealand, China, Austria, Denmark, Finland, France, Germany, Italy, Kuwait, Lebanon, Norway, Qatar, Russian Federation, Spain, Sweden, Switzerland, and United Kingdom.
  • Descripción del producto
    PROSTIVA RF Model 8929 Hand Piece is the delivery system component of the PROSTIVA RF Therapy System. || PROSTIVA RF Therapy Kit 8929, Sterilized using ethylene oxide, Manufacturer Medtronic Inc Minneapolis, MN 55432-5604, Manufactured at Medtronic, Inc Villalba, Puerto Rico. || Indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 cm3 and 50 cm3.
  • Manufacturer
  • Modelo / Serial
    Kit Lot Number: 215251, Hand Piece Lot Number: 13E005M.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Nationwide Distribution-including the states of AZ, CA, FL, GA, IL, KY, MA, MN, NC, NJ, NV, NY, OH, OR, TN, TX, and UT.
  • Descripción del producto
    Prostiva RF Therapy Hand Piece Kit, REF 8929, Rx only, sterilized using Ethylene Oxide. || The Prostiva RF Model 8929 Hand Piece Kit is the delivery system component of the Prostiva RF Therapy System. Prostiva RF Therapy delivers low-levels of radiofrequency (RF) energy directly into the hyperplastic tissue for the treatment of benign prostatic hyperplasia (BPH). The RF energy delivered by the system provides selective thermal ablation of prostate tissue to help reduce urethral || constriction and relieve BPH voiding symptoms. || The Prostiva RF Therapy System consists of the following major components: || " Radio Frequency Generator || " Hand Piece Kit || " Telescope || The Model 8929 Prostiva RF Therapy Hand Piece Kit contains the following single-use, disposable components: || " One sterile Hand Piece with radio || frequency cable (Model 8929) || " One sterile Tubing System (Model 6101) || " One non-sterile Return Electrode (Model 8934) || The hand piece has two insulated electrode needles that deliver the radio frequency energy interstitially to the prostate. These needles are covered by insulating shields that localize the lesions within the prostate, protect the urethra from thermal damage, and eliminate collateral thermal damage. The Hand Piece stainless steel (SS) tube is inserted into the urethral canal up to the prostate with the aid of a telescope for visualization during the therapy. The Hand Piece features two insulated Nitinol needles inside the SS tube. The needles can be deployed (extended), at approximately a 90-degree angle from the SS tube, to a preset depth of between 12 mm and 22 mm (at 2 mm intervals). The needles inside the SS tube are deployed into, or retracted from the prostate, using the handle lever. RF energy is delivered through the needles to the prostate tissue. The RF energy creates molecular friction eventually resulting in tissue heating and cell death (therapy intent).
  • Manufacturer
  • Modelo / Serial
    PROSTIVA RF Model 8930 Generator (Used for minimally invasive treatment for patients with lower urinary tract symptoms due to benign prostatic hyperplasia (BPH))ARTG Number: 137006
  • Manufacturer