Worldwide Distribution - USA (nationwide ) including the states of : AZ, NY, TN, MS, AR, WV, VA, VT, PA, DE, DC, OH, CO, NH, MI, CA, FL, CT, WA, MD, NM, RI, MA, TX, NC, IL, KY, WI and MI., and the countries of CANADA, SAUDI ARABIA and AUSTRALIA.
Descripción del producto
Medfusion¿ 4000 Syringe Infusion Pumps || Medfusion¿ 4000 Syringe Infusion Pumps are designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates. These pumps are indicated for use by healthcare professionals in the delivery of IV solutions, drug solutions, blood or blood products, and enteral feedings in continuous, intermittent, and body weight delivery modes.
GemStar docking station for use with specific GemStar infusion pumps.
The specific products potentially affected by the two problems are shown below:
Affected product (docking station list numbers)
Issue
Potential to occur in conjunction with products
13075-XX-07
13075-XX-09
Pump fails to power up
Gemstar phase 3 pump
(13000-XX, 13100-XX, 13150-XX)
Error code 11/003 on pump and infusion stops
Gemstar phase 4 pump
(13086-XX, 13087-XX, 13088-XX)
External battery pack
(13073-XX)
If you are unsure whether the Gemstar infusion pump is a phase 3 or a phase 4 contact Hospira for clarification.
There is no corrective action in place to resolve this issue. Hospira is offering the Sapphire pump, manufactured by Q Care, on a part exchange basis to replace Gemstar pumps as detailed in Medical Device Alert
MDA/2013/078
.