Device Recall Philips

  • Modelo / Serial
    This action affects certain customers who were shipped an affected Philips Intellivue Information Center unit between June 9, 2008 and May 31, 2009. Affected products have a PC serial number range between 2UA8300XYZ and 2UA84510QW. However, only a subset of these products are affected.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution - US (Nationwide) OUS: CANADA ARGENTINA AUSTRALIA AUSTRIA BAHRAIN BELGIUM BRAZIL CHILE CHINA COLOMBIA CZECH REPUBLIC FINLAND FRANCE GERMANY GREECE INDIA INDONESIA IRELAND ITALY JAPAN KENYA KOREA, REPUBLIC OF NETHERLANDS PHILIPPINES POLAND SPAIN SWEDEN SWITZERLAND THAILAND TURKEY UNITED KINGDOM
  • Descripción del producto
    M3150 Information Center Local Database; Models: 862249, 865001, 865003, 865004, 865007, 865009, 865011, 865012, 865091, 865092, 865093, 865094, 865095, 865096, 865098, 865099, 865102, 865104, 865105, 865125, 865127. || The intended use of the Information Center Software is to display physiologic waves, parameters and trends, formal data for strip recordings and printed reports and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors
  • Manufacturer
  • 1 Event

Fabricante

  • Dirección del fabricante
    Philips Electronics North America Corporation, 3000 Minuteman Rd, Andover MA 01810-1032
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA

229 dispositivos con un nombre similar

Más información acerca de la data acá

  • Modelo / Serial
    System Code #728243, Serial #: 7040, 7083, 7087, 7123, 7124, 7133, 7170, 7178, 7204, 7231, 7273, 7328, 7365, 7370, 7395, 7434, 7467, 7596, 7612, 7775, 7794, 7795, 7840, 7841, 7842, 7843, 7846, 7847, 7848, 7849, 7850, 7851, 7852, 7854, 7855, 7856, 7857, 7858, 7860, 7861, 7863, 7864, 7865, 7866, 7867, 7868, 7871, 7872, 7873, 7875, 7876, 7877, 7881, 7883, 7884, 7885, 7886, 7889, 7890, 7891, 7892, 7893, 7895, 7896, 7898, 7900, 7901, 7902, 7904, 7905, 7907, 7908, 7909, 7910, 7911, 7912, 7914, 7915, 7916, 7918, 7920, 7921, 7923, 7924, 7926, 7927, 7928, 7930, 7931, 7932, 7934, 7935, 7936, 7937, 7938, 7940, 7941, 7942, 7943, 7944 and 7945. System Code #728244, Serial #: 7121, 7271, 7764, 7844, 7845, 7853, 7859, 7862, 7869, 7870, 7874, 7878, 7879, 7880, 7882, 7887, 7888, 7894, 7897, 7899, 7903, 7906, 7913, 7917, 7919 and 7922. System Code #728246, Serial #: 3079, 3090, 3127, 3410, 3478, 3500, 3679, 3721, 5067, 5072, 5073, 5143, 5317, 5439, 5663, 5728, 5741, 5928, 5971, 6028, 6052, 6061, 6067, 6090, 6107, 6133, 6147, 6153, 6156, 6162, 6231, 6491, 6556, 6606, 6607, 6608, 6609, 6610, 6611, 6612, 6613, 6614, 6615, 6616, 6617, 6618, 6619, 6620, 6621, 6622, 6623, 6624, 6625, 6626, 6627, 6628, 6629, 6630, 6631, 6632, 6633, 6634, 6635, 6636, 6637, 6638, 6639, 6640, 6641, 6642, 6643, 6644, 6645, 6646, 6647, 6648, 6649, 6650, 6651, 6652, 6653, 6654, 6655, 6656, 6657, 6658, 6659, 6660, 6661, 6662, 6663, 6664, 6665, 6666, 6667, 6668, 6669, 6670, 6671, 6672, 6673, 6676, 6678, 6679, 6680, 6681, 6682, 6683, 6684, 6685, 6689, 6695, 6696, 6697, 6707, 6710, 6711, 6714, 6721, 30075, 40013, 50026, 50043, 50177, 50181, 50182, 50189, 50195, 50198, 50200 and 60009.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AR, AZ, CA, CO, CT, GA, HI, IL, IN, KS, MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR,TN, TX, UT, VA, VT and WV., and the countries of Australia, Austria, Bangladesh, Belgium, Brazil, Chile, China, Denmark, Egypt, France, Germany, Honduras, India, Indonesia, Italy, Kazakhstan, Latvia, Lebanon, Libyan Arab Jamahiriya, Mexico, Netherlands, Norway, Palestine, Philippines, Poland, Russia, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom and Vietnam.
  • Descripción del producto
    Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology and Brilliance CT 16 (Air) || Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology and Brilliance CT 16 (Air) scanners are whole body Computed Tomography X-ray systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories.
  • Manufacturer
  • Modelo / Serial
    Brilliance iCT: Model Number 728311 and 728306, Serial Number: 100103, 100019, 100023, 100170, 200047, 200013 200005 and 100087. Brilliance 64: Model Number 728231 and 728326, Seial Number: 4003, 9745, 9875, 90135, 90169, 95414, 300003, 300004, 300005 and 300010. Brilliance CT Big Bore, Model Number: 728243, Serial Number: 7006, 7060, 7154, 7298 and 7349.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution - US Distribution including the states of DC, IL, IN, MI, MO, NY, OH, OR, PA, TX and VT., and the countries of Australia, Belgium, China, Germany, India, Israel, Italy and The Netherlands.
  • Descripción del producto
    Brilliance iCT, Brilliance 64 and Brilliance Big Bore || The Brilliance iCT, Brilliance 64 and Brilliance Big Bore are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
  • Manufacturer
  • Modelo / Serial
    All Essenta DR x-ray Systems
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution - USA (Nationwide) including Puerto Rico and the countries of : Afghanistan, Argentina, Australia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hungary, India, Indonesia Iran,Italy, Jordan, Kazakstan, Kenya, Liberia, Lithuania, Morocco, Netherlands, New Zealand, Nigeria, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Thailand, Turkey, Ukraine, United Arab, United Kingdom and Viet nam.
  • Descripción del producto
    Philips Essenta DR, a Stationary x-ray system, X-ray system || Model: 712070 || The Essenta DR is a multifunctional stationary X-ray system, in which the swivel arm rotation, the vertical movement of the swivel arm and the SID adjustment are motorized. It is designed for radiographic examination of the recumbent, standing or seated patient and also for lateral exposures. It enables the operator to quickly and safely move the unit to all the pre-programmed basic positions at the press of a button, using the autopositioning feature. Fine positioning of the tube/collimator and detector on the patient is easy to carry out using the command.
  • Manufacturer
  • Modelo / Serial
    Software versions: F.02, R.02, 9.xx or any earlier versions   Addendum Models: M3532A, M3535A, M3536AM, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 M3535A- HeartStart MRx Defibrillator Monitor M3536A- HeartStart MRx Defibrillator Monitor EtC02 Options A04 A05 A06 All M3532A Upgrade
  • Clasificación del producto
  • Clase de dispositivo
    3
  • ¿Implante?
    No
  • Distribución
    US Nationwide and OUS.
  • Descripción del producto
    Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement || The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. The HeartStart MRx EtCO2 Option monitors carbon dioxide (CO2) and measures end-tidal carbon dioxide (EtCO2). The option provides an EtCO2 value and a CO2 waveform on the devices display.
  • Manufacturer
  • Modelo / Serial
    Serial numbers: US00558442 - US00558461 US00561519 US00558531 - US00558532 US00561578 US00558615 US00562454 US00558758 - US00558759 US00562504 - US00562527 US00560165 - US00560166 US00562640 - US00562641 US00560738 - US00560739 US00562832 US00560740 - US00560745 US00563125 US00560872 US00563231 - US00563243 US00560909 US00564032 US00561068 US00564177 US00561209 US00564409 US00561447 - US00561452 US00564793
  • Clasificación del producto
  • Clase de dispositivo
    3
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution - US Distribution including the states of IN, MS, TX and VA., and the countries of DENMARK, FRANCE, GERMANY, SPAIN and SWITZERLAND.
  • Descripción del producto
    Philips HeartStart MRx Monitor/Defibrillator || Models: M3536A Options A20 - A27, HeartStart MRx ALS || Monitor (Grey) || The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
  • Manufacturer
224 más