Device Recall Chemistry Wash1 REF 10680

  • Modelo / Serial
    Lots 16279, 17045 and 17215
  • Clasificación del producto
  • Clase de dispositivo
    1
  • ¿Implante?
    No
  • Distribución
    Worldwide and US Distribution.
  • Descripción del producto
    Wash Kit (REF 10680) Lots 16279, 17045 and 17215 || Medica corporation is marketing a wash solution Chemistry Wash1 , REF # 10680 for use with Medica s EasyRA clinical chemistry analyzer. The solution is used prior to a lipase (LIP) assay if an interferant assay (TRIG)is part of the work list. The major interferant for Lipase is the Triglycerides (TRIG) assay since one of the reagents contains lipase as an active ingredient.
  • Manufacturer
  • 1 Event

Fabricante

  • Dirección del fabricante
    Medica Corporation, 5 Oak Park Dr, Bedford MA 01730-1430
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA