Stryker Reflex Hybrid Screw Extractor

  • Modelo / Serial
    Ref # 48511905; all lots.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    Yes
  • Distribución
    Worldwide Distribution.
  • Descripción del producto
    Stryker Reflex Hybrid Screw Extractor; || Not Sterile: Manufactured by: Stryker Spine SAS, || Cestas, France; Distributed in the USA by: Stryker Spine, 2 Pearl Court, Allendale, NJ. || The threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the screw.
  • Manufacturer
  • 1 Event

Fabricante

  • Dirección del fabricante
    Stryker Spine, 2 Pearl Ct, Allendale NJ 07401-1611
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA

Dispositivo con un nombre similar

Más información acerca de la data acá

  • Modelo / Serial
    Ref # 48511905B, all lots.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    Yes
  • Distribución
    Worldwide Distribution.
  • Descripción del producto
    Stryker Reflex Hybrid Screw Extractor Inner Shaft; || Not Sterile; || Manufactured by: Stryker Spine SAS, Cestas, France; || Distributed in the US by: Stryker Spine, 2 Pearl Court, Allendale, NJ. || The threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the screw.
  • Manufacturer