Ingenuity CT

  • Modelo / Serial
    Ingenuity CT Model No. 728326 S/N:300004 30003 30005 30006 30007 30008 30009 30010 30011 30012 30013 30014 30015 30017 30018 30019 30020 30021 30022 30025 30026 30027 30028 30029 30030 30031 30032 30033 30034 30036 30037 30038 30039 30040 30041 30042 30043 30044 30045 30046 32083 300015 300033 300037 300038 300045 300054 300057 300066 300068 300072 300081 300082 300084 300089 300091 300107 300109 300120 300122 300141 300142 300149 300150 300156 300157 300207 300208 300209 300210 300211 30004-132107 52060 300003 300005 300010 300011 300012 300013 300014 300016 300017 300018 300019 300020 300021 300022 300024 300025 300027 300028 300029 300030 300031 300032 300034 300035 300036 300039 300040 300041 300042 300043 300044 300047 300049 300050 300051 300052 300053 300055 300056 300058 300059 300060 300061 300062 300063 300064 300065 300069 300070 300071 300073 300074 300075 300077 300078 300079 300080 300083 300086 300087 300088 300090 300092 300093 300094 300095 300096 300097 300098 300099 300100 300101 300102 300103 300104 300105 300106 300108 300110 300111 300112 300113 300114 300115 300116 300117 300118 300119 300121 300123 300124 300125 300126 300127 300128 300129 300130 300131 300132 300133 300135 300136 300137 300138 300139 300140 300143 300144 300145 300147 300148 300151 300158 300159 300160 300161 300162 300163 300164 300165 300166 300200 300201 300202 310058 310144 333057 336061 336084 336121 30047 52051 300213 300214 300216 300217 600023 600047
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Product was shipped to the following states: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WV & WY. Product was also shipped to the following countries: Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Cambodia, China, Colombia, Cuba, Estonia, Finland, France, Germany, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Philippines, Poland, Portugal, R¿union, Russian Federation, Senegal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Uzbekistan & Vietnam.
  • Descripción del producto
    Ingenuity CT Model No. 728326; To produce cross-sectional images of the body.
  • Manufacturer
  • 1 Event

Fabricante

  • Dirección del fabricante
    Philips Medical Systems (Cleveland) Inc, 595 Miner Rd, Cleveland OH 44143-2131
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA

187 dispositivos con un nombre similar

Más información acerca de la data acá

  • Modelo / Serial
    Model #728326, Serial Numbers: 300003, 300004, and 300005.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution - Nationwide (USA) including the states of: IN, MI, MO, NY, OH, and OR and the countries of Germany, Israel and Netherlands.
  • Descripción del producto
    Ingenuity CT (Computed Tomography) System, 510 (k) #K033326, Model #728326, Serial Numbers: 300003, 300004, and 300005. The recalled CT systems are equipped with Software Version 3.5 Beta 2. || Product Usage: The Brilliance Big Bore is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at difference angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
  • Manufacturer
  • Modelo / Serial
    The following Serial Numbers of this device are subject to recall: Serial Numbers: 300005, 300010, 300011, 300012, 300013, 300014, 300015, 300016, 300017, 300018, 300019, 300020 and 300021.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution - USA (nationwide) in the states of: LA, NY, OH, WV and the foreign countries of: Australia, Germany, Korea, India, Malaysiaa and New Zealand.
  • Descripción del producto
    Ingenuity CT (Computed Tomography) System, 510 (k) #K033326, Model #728326, Serial Numbers: 300005, 300010, 300011, 300012, 300013, 300014, 300015, 300016, 300017, 300018, 300019, 300020 and 300021. || Product Usage: || The Ingenuity CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
  • Manufacturer
  • Modelo / Serial
    The following Serial Numbers are subject to recall/software correction: 300003, 300004, 300005, and 300010.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide distribution: USA (nationwide) including states of: IN and OR; and countries of: Australia and Germany.
  • Descripción del producto
    Ingenuity CT (Computed Tomography) System, 510 (k) #K033326, Model #728326, Serial Numbers: 300003, 300004, 300005, and 300010. || Product Usage: The Ingenuity CT is a Computed Tomography X-Ray System intended to produce cross-a¿sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
  • Manufacturer
  • Modelo / Serial
    System Code # 728231, Serial # 91003, 95130, 95157, 95504 & 300030; System Code # 728326, Serial # 300010 & 300018.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution - USA including OH and the following countries: Australia, China, Belgium, Israel & Sweden.
  • Descripción del producto
    Whole body Computed Tomography X-Ray System (Brilliance 64 and Ingenuity CT), Brilliance 64 M/N 728231, Ingenuity CT M/N 728326, Dist. By Philips Medical Systems, Cleveland, OH
  • Manufacturer
  • Modelo / Serial
    System Code #728231, 4003, 4021, 9033, 9037, 9041, 9045, 9048, 9062, 9064, 9070, 9075, 9080, 9108, 9120, 9142, 9151, 9196, 9197, 9201, 9202, 9205, 9206, 9208, 9214-9218, 9462, 9508, 9531, 9534, 9535, 9544, 9554, 9561, 9584, 9603, 9606, 9608, 9613, 9627, 9630, 9657, 9668, 9671, 9672, 9695, 9706, 9708, 9719, 9728, 9734, 9735, 9741, 9745, 9756, 9789, 9819, 9832, 9838, 9839, 9840, 9859, 9874, 9875, 9891, 9898, 9904, 9918, 9926, 9952, 9964, 9965, 9968, 9975, 9997, 9999, 10000, 10008, 10015, 10041, 10052, 10054, 10092, 10095, 10100, 10103, 10114, 10125, 10145, 10151, 10152, 10154, 10170-10172, 10192, 10196, 10198, 10221, 10222, 10251, 10256, 10273, 10286, 10294, 10307, 10312, 10336, 10337, 10350, 10352, 10362, 10363, 10372, 10376, 10380, 10386, 10393, 10412, 10425, 10426, 10428, 10438, 10445, 10446, 10452, 10459, 10461, 10468, 10470, 10471, 10473, 10482, 10486, 10489, 10491, 10493, 10497, 10503, 10505, 10511-105531, 10533-10555, 10557-10566, 10568-10575, 10577-10604, 10606, 10608-10620, 10622-10624, 10628-10640, 29037, 29049, 29057, 64040, 90036, 90054, 90058, 90063, 90080, 90100, 90102, 90106, 90117, 90135, 90169, 90182, 91003, 95015, 95055, 95065, 95074, 95077, 95101, 95108, 95130, 95133, 95138, 95140, 95157, 95168, 95174, 95178, 95184, 95185, 95191, 95207, 95243, 95295, 95315, 95324, 95327, 95350, 95357, 95358, 95362, 95389, 95403, 95414, 95425, 95429, 95443, 95455, 95504, 95516, 95529, 95534, 95568, 95570, 95572, 95573, 95575, 95602, 95605, 95609, 95613, 95624, 95630, 95638, 95652, 95659, 95664, 95665, 95677, 95684, 95687, 95697, 95700-95728 & 300074.  System Code #: 728326, Serial #: 9511, 300003-300005, 300010-300022, 300024, 300025, 300027-300066 & 300068-300073.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution-USA (nationwide) including the states of AK, AR, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MT, ND, NE, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA and WV, and the countries of Australia, Bahamas, Bangladesh, Belgium, Bolivia, Brazil, Chile, China, Colombia, Croatia, Denmark, Finland, France, Georgia, Germany, Great Britain, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Laos, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Poland, Russia, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom and Vietnam.
  • Descripción del producto
    Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Systems. Brilliance 64 Model# 728231, Ingenuity CT Model # 728326, Distributed By: Philips Healthcare, Cleveland, OH. || The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.