Retiro De Equipo (Recall) de Access 2 Immunoassay Systems, UniCel DxC 600i SYNCHRON Access Clinical System An in vitro diagnostic device (IVD)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Beckman Coulter Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2015-RN-00143-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2015-03-13
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Beckman coulter is advising the users that it is releasing access 2 system software version 3.4.2 and access 2i (unicel dxc 600i system) software version 6.2.2 to address the reagent pack volume issues notification that was distributed august 13, 2010. the notification discussed the risks associated with assay reagent packs that were loaded incorrectly, or reagent packs that were inappropriately transferred between two systems (pack sharing). with these software versions, the new reagent pack monitoring feature enables the access 2 system pressure monitoring hardware to detect packs with insufficient volume, as well as packs that are improperly loaded or missing.
  • Acción
    Beckman Coulter is providing customer-installable 3.4.2 or 6.2.2 software disc along with an information packet that describes the new reagent pack monitoring feature to all customers in Australia. This action has been closed-out on 20/06/2016.