Retiro De Equipo (Recall) de Acor and Integrale primary femoral stem with modular neck

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Amplitude Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2017-RN-00497-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2017-05-09
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Amplitude would like to emphasize that patient selection is a key factor for the primary or revision implantation of the acor and integrale primary femoral stem with modular neck systems. patients with both a large lever arm and a heavy weight often cause severe stress on the modular joint.Reports of femoral neck fractures for certain combinations of modular necks and femoral heads have been received by the manufacturer, amplitude. subsequently, additional mechanical tests determining the fatigue strength of the proximal stem geometry have been conducted on the modular neck stems according to the recent standard iso 7206 -6: 2013 in order to determine which combinations of necks and heads should be subjected to usage restrictions.The surgical technique and the ifu have been updated with these usage restrictions regarding combination bans and admissible patient weight limits.
  • Acción
    Neosurgical is updating the IFU and Surgical Technique to highlight to surgeons the importance of observing the head/neck combination restrictions when using the device. Users are also advised to report any adverse reactions observed with the devices to Amplitude and/or to the Therapeutic Goods Administration (TGA).

Device

  • Modelo / Serial
    Acor and Integrale primary femoral stem with modular neckReference Numbers: 1-0106701, 1-0106702, 1-0106703, 1-0106704, 1-0191001, 1-0191002, 1-0191003 and 1-0191004Certificate Number: DV-2013-MC-04752-1ARTG Number: 210537