Retiro De Equipo (Recall) de Anti-dsDNA Kit. An in vitro diagnostic medical device (IVD).

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Immuno Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2014-RN-01362-1
  • Clase de Riesgo del Evento
    Class III
  • Fecha de inicio del evento
    2014-12-24
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Specific product codes for the anti-dsdna kit (used to test for antibodies to dsdna), were omitted in error from the packaging material. customers have received the correct products, however due to the omission of the product code on the outer box it is not clearly identifiable if they have received the 50 tests version (im77 lot c323026) or 100 tests version (im771 lot c323027). the kits can be differentiated by the lot numbers printed on the outer box. please note that the individual component labelling within the kit has not been affected.The tracer and ammonium sulphate solution lot numbers were omitted for these particular lots of product listed above.
  • Acción
    A set of labels (stickers) will be provided to customers to place over the existing directions that will correct the problem. The Customer Letter contains detailed instructions on where to place the stickers – including examples in the form of pictures. This action has been closed-out on 02/03/2016.

Device

  • Modelo / Serial
    Anti-dsDNA Kit. An in vitro diagnostic medical device (IVD).50 Test kit Batch Number:- C323026 Product Code IM77100 Test kit Batch Number:- C323027 Product Code IM771
  • Manufacturer

Manufacturer

  • Source
    DHTGA