Retiro De Equipo (Recall) de Application Instrument for Sternal ZIPFIX(Sternal ZIPFIX is used for closure of the sternum following sternotomy to stabilize the sternum and promote fusion)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Johnson & Johnson Medical Pty Ltd t/a DePuy Synthes.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2015-RN-00897-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2015-09-15
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    In the affected lots of the application instrument for sternal zipfix :· the end cap may loosen, thus reducing the tension applied to the implant. · the end cap may detach, allowing the tensioning spring to also become detached, making the instrument non-functional.If the end cap is loose, the maximum tension applied to the implant is reduced and may lead to insufficient sternal bone reduction. if the tension coil spring detaches completely from the zipfix application instrument while closing the sternum, it is possible that the spring or nut could fall into the thoracic cavity and go undetected. if the nut/spring is retained in the thoracic cavity, adverse tissue reaction may occur. no such occurrence has been reported to date.
  • Acción
    Customers are requested to immediately check their inventory to determine if the facility has any affected product and if so to quarantine those units prior to returning them to JJM for repair. This action has been closed-out on 05/09/2016.