Retiro De Equipo (Recall) de Artis Systems with 19" Live Display(fluoroscopic angiographic x-ray system)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Siemens Healthcare Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2016-RN-00924-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2016-07-11
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Due to an error in the 19" live display, image reproduction may fail in the examination room. the loss of image can only occur immediately after system start-up. this failure does not occur during normal operation or when the live image is displayed. only systems which use a live display (reference dshc1914-dc with part number 10656055) in the examination room are affected. live displays with other part numbers are not affected by this issue.
  • Acción
    Siemens is advising users that the system can continue to be used. A firmware upgrade will be implemented as a permanent correction.

Device

  • Modelo / Serial
    Artis Systems with 19" Live Display(fluoroscopic angiographic x-ray system)Reference Number: DSHC1914-DCPart Number: 10656055ARTG Number: 273952