Retiro De Equipo (Recall) de Artis Zeego Systems (Fluoroscopic diagnostic x-ray system for angiographic procedures) with SW revision VC14, VC20 & VC21 in conjunction with a specific technical configuration

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Siemens Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2012-RN-01165-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2012-11-30
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    In the course of siemens' product monitoring activities, a potential risk for patient or operators was determined during the operation of an artis zeego system with sw revision vc14, vc20 & vc21 in conjunction with a specific technical configuration which cannot be completely excluded. in case of an unlikely failure of brake, the c-arm may descent.This recall action was undertaken by siemens ltd before it was agreed with the tga.
  • Acción
    All the affected systems were checked by Siemens technical engineer for the potential for the failure mode to occur.

Device

  • Modelo / Serial
    Artis Zeego Systems (Fluoroscopic diagnostic x-ray system for angiographic procedures) with SW revision VC14, VC20 & VC21 in conjunction with a specific technical configuration ARTG Number: 102177