Retiro De Equipo (Recall) de Biograph mCT and Biograph mCT Flow

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Siemens Healthcare Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2016-RN-01650-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2016-12-23
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    An issue has been identified that the light marker windows used for the positioning laser and integrated into the front cover of the biograph mct or biograph mct flow system may loosen and potentially drop out. in this case, there is a risk that a person could touch the electrical or rotating parts of the gantry in the unlikely event that they reach through the window opening. this could result in a serious injury such as electrical shock or physical harm. there have not been any reports of complaints regarding this issue.
  • Acción
    Siemens is advising users to conduct a visual inspection of their Biograph mCT or Biograph mCT Flow and that in the event that the light market window is missing or loose, they should not use the Biograph mCT or Biograph mCT Flow and should immediately contact their local Siemens Customer Service Engineer for correction of the problem.

Device

  • Modelo / Serial
    Biograph mCT and Biograph mCT FlowCatalogue/Lot numbers: 10248668, 10248669, 10248670, 10248671, 10248672, 10248673, 10250743, 10250745, 10507786, 10528955, 10529159ARTG number: 144218
  • Manufacturer

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    DHTGA