Retiro De Equipo (Recall) de Blood Urea Nitrogen (BUN) Flex Reagent Cartridge (used with Dimension Vista System). An in vitro diagnostic medical device (IVD)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Siemens Healthcare Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2016-RN-00757-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2016-06-08
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    In february 2016, siemens healthcare diagnostics issued a letter, informing customers of well-to-well accuracy shifts for specific lots of dimension vista bun causing inaccurate patient and/or quality control results. preliminary investigation demonstrated loss of enzymatic activity of the gldh reagent wells made from a specific mold/cavity. siemens have since confirmed there is an additional cause of reduced gldh activity. · if calibration is performed using an unaffected well and samples are tested using an affected well, bun results may be falsely depressed by up to approximately 50% across the concentration range of 11 – 784 mg/dl. qc will detect the issue.· if calibration is performed using an affected well and samples are subsequently tested using an unaffected well, bun results may be falsely elevated by up to approximately 64% across the concentration range of 15 – 740 mg/dl. there may be in the delay in the follow-up for kidney disease.
  • Acción
    Siemens is providing work around instructions to users to follow in order to detect the affected reagent well sets. The actions are strongly recommended to minimise the usage of affected reagents wells. Additionally, users are advised to discard any reagent cartridges with cavity number B4. Siemens is not recommending a review of previously generated results however, a review due to this issue is at the discretion of the laboratory director. This action has been closed-out on 20/01/2017.

Device

  • Modelo / Serial
    Blood Urea Nitrogen (BUN) Flex Reagent Cartridge (used with Dimension Vista System). An in vitro diagnostic medical device (IVD)Catalogue Number: K1021Siemens Material Number: 10445159Lot Numbers: 16004AB, 16033AC, 16034AC, 16048AA, 16055AA, 16055AB, 16062AA, 16062AB and all future Vista BUN Flex reagent cartridges lots until further noticeARTG Number: 181689
  • Manufacturer

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    DHTGA