Retiro De Equipo (Recall) de Codman Certas Programmable Valves (Hydrocephalic valve) In Line Valve only with SIPHONGUARD Device, In Line Valve with SIPHONGUARD, Unitized Catheter and Accessories

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Johnson & Johnson Medical Pty Ltd T/A Depuy Australia.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2013-RN-00473-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2013-05-17
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Johnson & johnson testing has shown that, in a small percentage of valves, the following conditions may exist:1. the mri resistance feature may not always operate properly. if this occurs, it could potentially result in an unintended change in the valve setting due to exposure to an mri procedure or other magnetic field. based on reported complaints, this issue may have been a factor in up to 0.06% of units sold.2. the programming mechanism may not always operate properly. if this occurs, it could potentially lead to an inability to modify the operating pressure of the valve with the hand-held programming tools, also referred to as the codman certas therapy management system (tms). based on reported complaints, this issue may have been a factor in up to 0.4% of units sold.
  • Acción
    JJM is recalling the affected units that have not been implanted. For patients that have already been implanted with the device, JJM is providing additional instructions to clinicians to identify whether the implanted device has been affected and what actions to undertake if the device has been affected. For more details, please see http://www.tga.gov.au/safety/alerts-device-codman-certas-valves-130527.htm . This action has been closed-out on 29/01/2016.