Retiro De Equipo (Recall) de Eluvia 150mm and Innova 200mm Stent Systems

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Boston Scientific Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2017-RN-01391-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2017-11-07
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Boston scientific is initiating a removal from the market of the150mm eluvia drug-eluting vascular stent system and the 200mm innova self-expanding stent system due to elevated complaint rates for partial stent deployment.There have been 190 complaints of partial deployment. partial deployment occurs when the stent is unable to be fully released from the delivery system. part of the stent can become anchored in the vessel while the rest of the stent remains within the delivery system. the most common reported injury has been additional medical or minor surgical intervention, vessel trauma or prolongation of the implant procedure. however, boston scientific has received reports in 14 cases of major surgery to retrieve the stent/delivery system or to correct vascular compromise.
  • Acción
    Boston Scientific is advising users to inspect stock and quarantine the affected products for return for credit.

Device

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Comentario del fabricante
    “We take a patient-first approach to assessing the applicability of every recall and communicate to regulatory bodies in all geographies where the recalled device is sold,” Boston Scientific said in a statement to ICIJ. “We have coordinated several recalls across many countries in a timely manner,” the company said, adding that it complies with all national laws, which can often vary and require different processes for reporting information or taking action on recalls. The company said it uses a rigorous and uniform process to take action on recalls and that “when we initiate a field action (e.g. recall, safety alert), every customer who has received an affected product receives a communication that includes a letter for the physician.”
  • Source
    DHTGA