Retiro De Equipo (Recall) de Eon Wall Charger Model 3701 and Eon Portable Charger

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por St Jude Medical Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2015-RN-00460-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2015-05-29
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Some patients have reported experiencing an uncomfortable temperature increase at the ipg implant site during charging, with a small number of patients reporting burns at the implant site. this matter is a follow-up to a hazard alert letter issued by st. jude medical (sjm) in august 2012 (tga reference: rc-2012-rn-00759-3). in the august 2012 letter, st. jude medical informed customers that they would be implementing design improvements to the charger to address the possibility of increased energy dissipation during charging. the new eon charging system model 3726 reduces the potential for uncomfortable temperature increases that may be felt near the spinal cord stimulator while recharging by automatically turning the energy it delivers on and off to reduce the rise in temperature of the spinal cord stimulator and of the charger wand during charging.
  • Acción
    St. Jude Medical is recommending to implanting physicians for the prompt replacement of the charger for patients who have reported excessive heating during charging. For other patients, replacement of the charger is recommended at the next follow up. Upon replacement, proper charging function of the Eon Charging System Model 3726 should be confirmed. Until such time that patients have had their chargers replaced, patients should be advised to continue to use their current Eon Mini Charging System following the precautions provided in the August 2012 letter. This action has been closed-out on 09/08/2016.

Device

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Comentario del fabricante
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    DHTGA