Retiro De Equipo (Recall) de Graseby 3300 Pump and Graseby 3000 Series Accessories

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Smiths Medical Australasia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2017-RN-01146-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2017-09-06
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Smiths medical has become aware that in certain cases, the lock and key supplied with graseby 3300 pumps and accessories does not match the standard graseby 3300 key. affected products were either serviced or manufactured between 30 oct 2015 and 06 apr 2016.In affected cases, the key provided will operate the lock, but cannot be shared between affected pumps and other graseby 3300 pumps as it should. the key is needed to unlock the pump’s security cover to allow programming of infusion, changing of program settings, restarting infusion or allow clinician overrides. it is also required for operation of the ratcheting type pole clamp (described as the “locking” type): the lock must be operated to remove the pump from the pole and to reposition the pump on the pole.A mismatch between the lock and key has the potential to delay the delivery of pain medication and lead to an increase in pain. to date, smiths medical has not received any reports of serious injuries related to this issue.
  • Acción
    1. Complete and return the supplied Form even if you do not have any affected devices in your possession. 2. Locate all keys associated with GRASEBY pumps and locking accessories and visually inspect keys to ensure keys are labelled “95001”. Collect and isolate any keys not labelled “95001”. 3. Ensure all personnel with Graseby keys are using key “95001”. 4. Using the “95001” key, unlock and relock the locking mechanism on all GRASEBY pumps or locking accessories. 5. If the “95001” key does not successfully operate the locking mechanism, or if you do not have a “95001” key, discontinue use of the pump. 6. Return affected keys to Smiths Medical Australasia. A representative will contact you about scheduling replacement of the affected key. 7. Notify anyone to whom you have distributed this product.

Device

  • Modelo / Serial
    Graseby 3300 Pump and Graseby 3000 Series AccessoriesGraseby 3300 PumpsModel Number: 0128-0001Lock AssemblyModel Number: 0128-0025Pole ClampModel Number: 0128-0135 and 0128-0136Security Cover KitModel Number: 0128-0451ARTG Number: 122405