Retiro De Equipo (Recall) de Hamilton Medical MR1 Ventilator

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por The Trustee for the Taleb Family Trust T/A Taleb Medical and Vetequip Australia.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2016-RN-01133-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2016-08-29
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    An issue has been identified where the oxygen tubing and the oxygen connector of the hamilton-mr1 became loose during the preparation for ventilation. in the case of loose oxygen tubing during operation, the ventilation is continued and the “oxygen supply failed” alarm is triggered if the oxygen setting is above 21%. the oxygen connector may hurt the patient and the operator due to the high pressure of the oxygen and uncontrolled movement of the tubing. there have been no injuries as a result of this issue.
  • Acción
    Taleb Medical is advising users to check the oxygen connector screws for tightness as an immediate measure. Tighten with a screwdriver if necessary. Do not over tighten the screws as this can damage the thread. If the oxygen connector cannot be tightened or if the screw-hole threads are damaged, do not use the ventilator. A new O2 mixer mounting plate will be installed as a permanent correction.

Device