Retiro De Equipo (Recall) de Infant Nasal CPAP Prongs and Bubble CPAP Starter Kits

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Fisher & Paykel Healthcare Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2014-RN-00437-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2014-04-14
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Fph is issuing this recall notice because it has received customer reports relating to certain lot numbers of certain models of nasal cpap prongs detaching from the fph flexitrunk nasal tubing, especially in the presence of mucous and/or moisture. in the event of the affected prongs detaching from the nasal tubing, therapy is likely to be interrupted. this may lead to a patient becoming hypoxic.
  • Acción
    FPH is requesting their customers to identify any units form the affected lot and destroy them. An FPH Representative can assist in sourcing a replacement product.