Retiro De Equipo (Recall) de Intro-Flex Automatic Hemostasis Valve Introducer

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Edwards Lifesciences Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2017-RN-01488-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2017-12-05
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    This recall involves an introducer sheath that was found to be out of specification. if the sheath inner diameter is too small, the clinician may have difficulty inserting the dilator or catheter into the introducer. this will cause a procedural delay as the clinician will need to change out the device over a guidewire with a minimal delay in treatment.There have been no injuries as a result of this issue.
  • Acción
    Edwards is advising users to inspect stock and quarantine any remaining units of the affected lot. Affected units can be returned to Edwards.

Device

  • Modelo / Serial
    Intro-Flex Automatic Hemostasis Valve IntroducerModel Number: I300F85Lot Number: 60377828ARTG Number: 146688