Retiro De Equipo (Recall) de Masimo uSpO2 and HPLP Oximetry Cables

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Philips Electronics Australia Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2015-RN-00542-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2015-06-26
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    A small number of oximetry cables have been identified with crossed-internal wires.The effect on performance when these wires are crossed is that when a low spo2 value is measured, a high spo2 value would be displayed on the monitoring device. additionally, in the same oximetry cable, when a high spo2 value is measured, a low spo2 value would be displayed on the monitoring device.
  • Acción
    Philips is providing instructions for hospitals to verify the performance of the oximetry cables. If the cables are not working as expected they are to be returned to Philips for replacement or refund. This action has been closed-out on 06/09/2016.