Retiro De Equipo (Recall) de ME DPHC 0001 deltastream HC, 230V (Cardiopulmonary bypass system heating/cooling unit)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Lovell Surgical Supplies International Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2016-RN-00362-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2016-04-07
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    There have been reported cases of nontuberculous endocarditis caused by mycobacteria or infections after cardiac surgery, in which hypothermia or hyperthermia devices were used. the manufacturer is updating the cleaning procedure of the device as a precautionary measure. updated ifus are expected to be available by the end of june 2016.
  • Acción
    The manufacturer is advising users to follow the following updated cleaning procedures: · Clean the device every 14 days with Sanosil solution (3% H202). Since there may be an excess pressure during the cleaning process the tank lid must remain open during the cleaning process Furthermore, it is recommended that the water is changed daily and after each patient. · Change least once a year, the hose assembly. · All relevant hypothermia and hyperthermia devices should be operated in a separate room from the operating theatre. If this is not possible, another suitable measure should be found to ensure the risk of infection is effectively minimised. For example, by increasing the disinfection frequency of the surfaces of the devices and changing gloves after cleaning the equipment. · Perform regular checks for contamination of equipment. The manufacturer recommends monthly intervals to determine the quality of the water.