Retiro De Equipo (Recall) de MGuard Prime Coronary Stent System Embolic Protective Stent (EPS)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Emergo Asia Pacific Pty Ltd T/a Emergo Australia.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2014-RN-00545-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2014-05-20
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    The manufacturer, inspiremd, has initiated this recall action due to recent complaints of mguard prime eps dislodgements. these complaints have primarily occurred during the preparation of the mguard prime eps, upon removal of the protective sleeve, or during withdrawal of the mguard prime eps into the guide catheter during the procedure. these complaints have not resulted in any patient injury.
  • Acción
    Upon getting appropriate regulatory approval, InspireMD intends to perform a manufacturing enhancement to all unexpired units of the MGuard Prime EPS systems to improve stent retention and product performance.