Retiro De Equipo (Recall) de ML 301 Examination Lamp

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Medsell Sales Pty Ltd T/A Denyers International.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2013-RN-00236-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2013-03-20
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    There are reports that after an average usage period of seven years fatigue failure of the plastic joint may occur. the manufacturer has become aware that a joint on the spring arm may break either during use or when the spring arm is not in operation. if the joint breaks, the spring arm falls suddenly and without hindrance. in such a case there is a high risk of injury to the patient and other persons in the immediate vicinity.This action was undertaken without consulting the tga.
  • Acción
    Denyers International inspected all units to determine which units required the correction. All units were identified as not affected by the issue.

Device