Retiro De Equipo (Recall) de MSK 1.5T Extreme and Optima MR430s (MRI Systems)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por GE Healthcare Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2016-RN-00652-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2016-05-13
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    A site reported a magnet quenching with subsequent venting of the cryogen gas into the magnet room. cryogen gas that vents into the magnet room could cause critical deprivation of oxygen to users and patients. there have been no injuries reported as a result of this issue. at this site, a magnet heater probe connection became disconnected. this resulted in ice build-up within the magnet blocking the cryogen vent. since the vent was blocked by ice, the later magnet quench caused gas to be vented into the magnet room.
  • Acción
    GE Healthcare is advising users to notify GE Service Engineer if the Magnet Monitoring Unit (MMU) detects any abnormal or out of specification reading. If the Magnet Monitoring Unit (MMU) detects an error it will display a pop-up window on the system monitor which will be seen by the operator. MMU reported errors will be in the form of a CAS7XX error (XX being the specific error) with a brief description. If no Magnet Monitoring Unit errors are reported, there are no action is required. Users can continue normal use of the affected products. GE will be correcting all affected units. This action has been closed out on 13 Jun 2017.