Retiro De Equipo (Recall) de OASYS Midline Occipital Plate(Intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput –T3))

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Stryker Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2013-RN-00670-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2013-07-03
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Stryker spine has received reports from the field indicating post-operative failure of the pin that connects the tulip head to the plate body. the pin can fail when:- an excess acute load is applied to the pin causing the pin connecting the tulip to the plate to fracture; or- an excess repetitive load is applied to the pin causing the pin connecting the tulip to the plate to fracture (fatigue fracture).
  • Acción
    The manufacturer, Stryker Spine is currently undertaking an investigation regarding the complaints received. In order to minimise any potential risk to patients while the investigation is ongoing, Stryker Spine recommends routine clinical and radiographic post-operative evaluation for cervical fusion patients implanted with the device. However, should the patient report or exhibit any change in symptoms such as increased pain, weakness, or numbness, more urgent evaluation should be completed. For patients who have had a revision surgery, Stryker Spine recommends routine post-operative care for revised cervical spine fusion patients. For more information, please see http://www.tga.gov.au/safety/alerts-device-oasys-midline-plate-130722.htm .

Device

  • Modelo / Serial
    OASYS Midline Occipital Plate(Intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput –T3))Catalogue Numbers: 48571044, 48571045, 48571046, 48571047 and 48571048All lot numbers are affectedARTG Number: 179439
  • Manufacturer

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    DHTGA