Retiro De Equipo (Recall) de OPTEASE Retrievable Vena Cava Filter

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Johnson & Johnson Medical Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2013-RN-01088-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2013-10-22
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    A printing error has been identified on one unit of optease retrievable vena cava filter. the orientation arrow for the femoral approach was printed in the incorrect direction. the error resulted in the filter being implanted upside down, requiring an additional percutaneous procedure to retrieve the filter. all lots of the cordis optease retrievable vena cava filter are being removed, since it cannot be absolutely determined that no other similar printing errors occurred.
  • Acción
    Customers are requested to quarantine the affected product prior to returning to JJM.

Device

Manufacturer