Retiro De Equipo (Recall) de Paratherm Heater Cooler Unit

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Australian Blood Management Unit Trust.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2018-RN-00646-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2018-05-24
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Updated instructions for use (ifu) and cleaning guidelines for the paratherm heater cooler unit have been released.There was a safety alert communicated in april 2017 which referenced the interim cleaning and disinfecting guidelines. during the last 12 months, validation efforts have been undertaken by the manufacturer chalice medical to support the use of a hydrogen peroxide based disinfection agent. this disinfection agent has been subject to review and subsequent inclusion in the australian register of therapeutic goods (amity hp 75, artg 302115, supplied by lovell surgical supplies international as the local sponsor).
  • Acción
    Amity HP 75 is to be made commercially available in Australia in June 2018. Stock is currently being manufactured and shipped to Australia. Once available in Australia, Australian Blood Management Unit Trust will be able to facilitate supply of this disinfection agent through Lovell Surgical Supplies International.

Device

Manufacturer