Retiro De Equipo (Recall) de Patient Data Module (PDM) used with B850, B650, B450, Solar 8000i, MacLab / ComboLab (Multifunction patient monitoring)All PDMs produced from April 2007 through June 2010

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por GE Healthcare Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2013-RN-01255-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2013-11-27
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Fluid can pass through the seals in the top cover and come into contact with energised contacts in the circuit board. this can result in heat buildup, smoke, and discoloration. this heat buildup could potentially result in thermal injury to a patient or a caregiver.
  • Acción
    GE Healthcare is advising users to inspect the top cover for signs of heat buildup or discolouration. Remove from use if any evidence of damage is observed. GE is advising users to refrain from spraying fluid onto device. A new cover will be installed on all affected devices as a permanent fix when available. This action has been closed-out on 11/02/2016.

Device

  • Modelo / Serial
    Patient Data Module (PDM) used with B850, B650, B450, Solar 8000i, MacLab / ComboLab (Multifunction patient monitoring)All PDMs produced from April 2007 through June 2010Serial Numbers: SA30713XXXXXX to SA31026XXXXXXARTG Number: 118982