Retiro De Equipo (Recall) de PCA LifeCare Infusion System

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Hospira Pty Limited.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2013-RN-00292-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2013-04-19
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • Notas adicionales en la data
  • Causa
    Hospira has received reports of pca pumps not detecting distal occlusions, one of which resulted in serious injury. this issue is caused by wear and tear on the half nut which prevents it from properly detecting the pressure build-up associated with a distal occlusion.
  • Acción
    Hospira is recalling the affected pumps. The pumps have not been supplied since 2006.

Device

  • Modelo / Serial
    PCA LifeCare Infusion System Catalogue number: 0F7822703Number on pump: 0F782ARTG number: 138109