Retiro De Equipo (Recall) de Per-Q-Cath PICCs

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Bard Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2015-RN-00480-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2015-06-04
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    During an internal labelling review, bard identified that the product gauge size (ga) is incorrect on certain per-q-cath products. the product gauge (ga) size is provided in two locations: on the label and in some cases on the catheter hub. this issue has been identified as a conversion problem (french to gauge size). no adverse reaction reports or other indication of injuries or diseases have been reported relating to this problem.
  • Acción
    Bard is providing the correct Gauge measurement to hospitals and healthcare professionals. This action has been closed-out on 04/08/2016.

Device