Retiro De Equipo (Recall) de QuikFlap Neuro Implants

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Stryker Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2018-RN-00682-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2018-06-18
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    The current packaging and ifu indicate a five-year shelf life. during long-term product testing, it was discovered that after a three-year term, there is a potential for the external packaging (sterile barrier) to become compromised. such compromised packaging results in the loss of the vacuum seal allowing for contamination of the inside peel pouch. a breach of the outer peel pouch has only been observed with packages at shelf life end and after excessive transportation testing.To date, medtronic have not received any complaints as a result of this issue.
  • Acción
    Stryker is advising customers to quarantine affected stock to prevent further use. Stryker representatives will contact customers to help coordinate the return of any affected product. In the event this device has been recently implanted, surgeons are advised to monitor for signs of infection as per their normal processes.

Device

  • Modelo / Serial
    QuikFlap Neuro ImplantsMultiple Item NumbersMultiple Lot Numbers ARTG Number: 295882(Stryker Australia - Cranofacial fixation plate, non-biodegradable)
  • Manufacturer

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    DHTGA