Retiro De Equipo (Recall) de Rainbow Reusable Sensors for rainbow SET Devices with SpCO, SpO2, and SpMet. (used with pulse oximeter for the measurement of parameters such as arterial oxygen).

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Masimo Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2015-RN-01074-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2015-11-09
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Masimo has identified specific lots of sensors that were manufactured with incompatible configurations. this could result in sensors that will either not provide readings for the spco and spmet parameters or inaccurate readings. the specific behaviour will depend on the firmware version of the technology board in the device. spco measurements, when displayed, could be incorrectly elevated throughout the measurement range. spmet measurements above 6% could be affected by a negative bias of 1.5%, on average.
  • Acción
    Masimo is advising users to inspect stock and quarantine any affected units for replacement with unaffected stock. This action has been closed-out on 18/08/2016.

Device

  • Modelo / Serial
    Rainbow Reusable Sensors for rainbow SET Devices with SpCO, SpO2, and SpMet. (used with pulse oximeter for the measurement of parameters such as arterial oxygen). Reference: 2696 Lot Number: 15HA6Sensor Reference: 24087-A15G760 , 24087-A15H754ARTG Number: 157503