Retiro De Equipo (Recall) de Resuscitation Systems with blender field upgrade installed in GE Giraffe and Panda Warmers (infant radiant warmer)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por GE Healthcare Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2012-RN-01081-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2012-10-23
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Ge healthcare has recently become aware of a potential safety issue wherein the oxygen and air wall inlet fittings and/or labels on the back panel, were reversed during assembly. as a result, this may potentially have reversed air/oxygen mixer concentrations; for example, a setting of 100% oxygen could have an output of 21% oxygen, and vice versa. the settings of the blender knob will no longer be accurate.
  • Acción
    If the affected unit is in use with a patient, GE is asking customers to disconnect the wall supply gas and switch to tank supply. If the affected unit is not currently in use, GE is asking customers to take the unit out of service. A GE Field Engineer will be deployed to the affected site to complete the required repair.