Retiro De Equipo (Recall) de Sensei X Robotic Catheter System

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por St Jude Medical Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2014-RN-01325-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2014-12-19
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    There is a potential for the set-up joint sub-assembly to become loose relative to the operating table. should this issue occur during clinical use, it may potentially result in uncontrolled catheter motion.Uncontrolled catheter motion could result in cardiac tamponade or perforation of the heart.
  • Acción
    The Customer is requested to ensure that the RCM is secure and tightly connected to the patient table before starting a procedure and check that all clamps are properly tightened. In addition, ensure that the RCM does not come into contact with other laboratory equipment, as this may produce forces capable of dislodging the RCM. In the event the clamp is found to be loose, customers are advised not use the system and to contact St.Jude Medical immediately. This action has been closed-out on 14/07/2016.

Device

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Comentario del fabricante
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.