Retiro De Equipo (Recall) de Sorin 1T & 3T Heater Cooler (Perfusion system)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Cellplex Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2015-RN-00511-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2015-06-16
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Sorin has become aware that the actual disinfection practices and the water maintenance that some users have been performing are not always conducted according to our instructions for use. without vigilant performance of the disinfection and maintenance procedures per the instructions for use, organisms can multiply in a heater cooler device and potentially form biofilm.
  • Acción
    Sorin Group is providing end users with additional instructions for cases where the heater cooler has not been maintained according to the Instructions for Use. The IFU has also been updated to include the additional cleaning instructions. This action has been closed-out on 14/07/2016.

Device

  • Modelo / Serial
    Sorin 1T & 3T Heater Cooler (Perfusion system)All serial numbers starting with 16SXxxxx are affectedARTG Number: 194514
  • Clasificación del producto
  • Manufacturer

Manufacturer