Retiro De Equipo (Recall) de Sorin Group perfusion system – Heater Cooler 3T devices

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por LivaNova Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2017-RN-01368-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2017-10-27
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Livanova is issuing a recall for product correction to inform users about the retrofit of their heater cooler 3t devices to implement a vacuum and sealing upgrade. this modification to the internal design of the 3t heater-cooler device will collect any aerosols that are potentially contaminated with nontuberculous mycobacterium (ntm), preventing those aerosols from entering into the sterile operating field.Livanova is also notifying that the 3t device will be suspended from the australian register of therapeutic goods (artg) beginning 8 november 2017 for a period of up to 6 months. the reason for this suspension is that the minncare disinfectant is not listed on the artg with the same classification as the 3t device. however, during this period, livanova will be able to supply loan units on an as-needed, no-charge basis to ensure continuity of clinical practice.
  • Acción
    LivaNova has completed the development of a TGA approved correction that mitigates against this potential contamination by preventing dispersion of aerosols produced by the heater-cooler in the operating room. This correction will be implemented by an onsite enhancement of units currently in use by LivaNova service personnel or a local agent. Although this issue is less likely when the 3T device’s water system is maintained according to the current IFU, this product correction will mitigate against the reasonably foreseeable hazard of patient infection which may be potentiated by poor maintenance practices.