Retiro De Equipo (Recall) de Synthesis, Avant & Primo2x Families Adult Oxygenator Mounting Brackets (used with Sorin 3T Heater-Cooler devices)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por LivaNova Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2016-RN-00886-1
  • Clase de Riesgo del Evento
    Class II
  • Fecha de inicio del evento
    2016-07-13
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Livanova (sorin) have found that certain oxygenator mounting brackets that include tubing that connects the heater-cooler 3t water circuit to the oxygenator could be a source of re-contamination (with non-tuberculous mycobacteria) of a clean heater-cooler 3t’s water circuit. this tubing is not subjected to disinfection as per the ifu for the heater-cooler. if the heater-cooler 3t user failed to clean and disinfect the heater-cooler 3t resulting in contamination of the heater-cooler 3t and at some point later used oxygenator mounting bracket, the water from the contaminated heater-cooler 3t could cross-contaminate the oxygenator mounting bracket’s internal tubing in which biofilm could develop. if a non-contaminated heater-cooler 3t was later connected to an oxygenator mounting bracket that had been previously cross contaminated, the oxygenator mounting bracket could contaminate the clean heater-cooler 3t.
  • Acción
    LivaNova will replace the affected bracket and is providing advice to users on how to dispose of the affected Oxygenator bracket. The existing brackets can be used until a replacement unit can be provided.

Device

  • Modelo / Serial
    Synthesis, Avant & Primo2x Families Adult Oxygenator Mounting Brackets (used with Sorin 3T Heater-Cooler devices)D632 Adult Open System Bracket Catalogue Number: 05046Adult Oxy module Avant Bracket Catalogue Number: 050128Synthesis “C” BracketCatalogue Number: 050350Synergy Bracket Catalogue Number: 050131D626 Avant Holder Catalogue Number: 05040D630 Avant Holder Catalogue Number: 05041Synthesis Oxygenator Holder Catalogue Number: 050127ARTG # 194514
  • Manufacturer

Manufacturer