Retiro De Equipo (Recall) de Tenor without scale (Mobile patient transfer lift)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Huntleigh Healthcare Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2014-RN-00566-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2014-05-20
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    During use, the bolt in the connection between the spreader bar and the lifting arm jig is subjected to mechanical forces, which may lead to breakage. this is due to the design of the connection between spreader bar and lifting arm jig, which in long term use can cause metal fatigue in this part and the formation of micro-cracks in the thread of the bolt. this issue is only related to tenor devices without scales.
  • Acción
    ArjoHuntleigh will arrange for the replacement of the old spreader bar with a new spreader with scales. This action has been closed-out on 12/02/2016.

Device

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    DHTGA