Retiro De Equipo (Recall) de Ultrafilter U9000 (Used on Gambro Haemodialysis Delivery Systems.)

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por Gambro Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2014-RN-00860-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2014-08-05
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Gambro pty ltd is advising users of the potential for external fluid leakage from the ultrafilter u9000 due to multiple disinfections using a sodium carbonate-based solution. intensive use of sodium carbonate based disinfection programs over several months may cause degradation of the ultrafilter housing leading to an external fluid leakage at the position of the welding seal between the filter housing and the header cap. if the leakage is slow developing, there is a remote probability that it will not be noticed until the dialysis treatment has started.
  • Acción
    If a leakage is observed during treatment, the users are advised to immediately stop the treatment and change the U9000 Ultrafilter to avoid excessive fluid removal from the patient. Gambro is advising that the use of sodium carbonate-based disinfection programs has been broadly validated for the U9000 Ultrafilter. However, to minimise the risk of an external fluid leak, Gambro recommends adhering to the following instructions, when using sodium carbonate based disinfection programs: 1. Do not exceed the recommended disinfectant concentration given in the Operator’s Manual. 2. Do not use more than eight sodium carbonate disinfection cycles over the lifetime of the U9000. 3. Do not use sodium carbonate disinfection cycles more than once a week. 4. Include daily disinfection with citric acid in the disinfection protocol. 5. Avoid periods of machine inactivity (e.g. weekend storage) directly after disinfection with sodium carbonate based agents. 6. The recommended process of combined “Heat + CleanCart-A” disinfection shall be followed by a “Heat + CleanCart-C” disinfection on the same day and ideally be performed in the middle of the working week. This action has been closed-out on 03/06/2016.

Device

  • Modelo / Serial
    Ultrafilter U9000 (Used on Gambro Haemodialysis Delivery Systems.)ARTG Number: 131741
  • Manufacturer

Manufacturer