Retiro De Equipo (Recall) de Zenith Alpha Thoracic Endovascular Graft

Según Department of Health, Therapeutic Goods Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en Australia que fue producido por William A Cook Australia Pty Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    RC-2017-RN-00824-1
  • Clase de Riesgo del Evento
    Class I
  • Fecha de inicio del evento
    2017-06-28
  • País del evento
  • Fuente del evento
    DHTGA
  • URL de la fuente del evento
  • Notas / Alertas
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notas adicionales en la data
  • Causa
    Cook medical previously advised customers in march 2017 (rc-2017-rn-00419-1) about an update to the instructions for use (ifu) for the zenith alpha thoracic endovascular graft.Since then, cook medical has received additional complaints for the same issue. although the complaints were from patients treated prior to the march notification, there is now an additional correction of the (ifu) and the removal of specific sizes from the market.The correction to the ifu removes the indication for use in blunt thoracic aortic injury (btai), and a warning has been added to describe the thrombus risk that has been observed when the device is used to treat btai. patients already treated with this product for the btai indication should be followed up according to the current ifu, with considerations outlined in cook medical’s correspondence in march 2017.Potential adverse events that may occur if these devices were used for btai include death, paraplegia, and/or surgical intervention.
  • Acción
    Cook Medical is requesting that Customers: 1. Share this notice with others in their organisation who either use this device or follow patients treated with this device. 2. Follow the IFU corrections as provided. 3. Examine inventory and immediately quarantine affected products. 4. Complete and return the required Acknowledgement and Receipt Form, in addition to any affected recalled products to Cook Medical. Return Goods Authorisation Number: QCR-75 Cook Medical 61 Brandl Street Eight Mile Plains QLD 4113 5. Maintain a copy of this notice for their records. Upon availability of the corrected IFU, a Cook Medical Sales Representative will personally follow-up and provide corrected IFUs for inventory.

Device

  • Modelo / Serial
    Zenith Alpha Thoracic Endovascular GraftCatalogue Numbers: ZTA-D-/-ZTA-DE-/-ZTA-P-/-ZTA-PT-/- Multiple Catalogue Identifier and Global Product NumbersAll LotsARTG Numbers: 222505, 222526, 222533, 222525
  • Clasificación del producto
  • Manufacturer

Manufacturer