Alerta De Seguridad para ACCESS THYROGLOBULIN CALIBRATORS, Registration No. 10033120496, Lots 325451, 327168 and 330203

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por BECKMAN COULTER DO BRASIL COM. E IMP. DE PROD. DE LAB. LTDA.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1367
  • Fecha
    2014-02-24
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    This problem can result in high patient and quality control results. The false control results and the patient's thyroglobulin may lead to an additional test. Based on the internal test, the observed increase in results ranged from 11% to 18% when the calibrators were reconstituted with 2.0 mL of distilled water and allowed to stand for 30 minutes prior to use. All levels of control and patient outcomes were affected. The percentage increase in response observed depends on the reconstitution time before use for the affected lots. A reconstitution time of 1 hour and 40 minutes or more using the affected batches provided appropriate recovery of control and patient outcomes. ///// UPDATE - 4/3/2014 - The company forwarded a response letter to all customers who received the aforementioned lots informing them of the necessary actions and forwarded an uncommitted lot. If any service still has lots quoted in this alert, please contact the company.
  • Causa
    Beckman coulter of brazil received beckman coulter inc., an important product notice stating that the 30 minute reconstitution time is inadequate to fully reconstitute batches 324753, 325451, 327168 and 330203 of the access thyroglobulin calibrator.
  • Acción
    The company directs customers to: i- Discard all Access Thyroglobulin Calibrator kits listed above. Determine when batches of calibrator number 324753, 325451, 327168, and 330203 were calibrated in your system and evaluate all results generated with the Access Thyroglobulin Calibrator to determine if the control and / or patient results were higher than expected or did not match the patient's clinical status. iii - Under the direction of the Laboratory Director, notify physicians that it is possible that the thyroglobulin sample results reported by your laboratory are affected by this problem.