Alerta De Seguridad para Accessory Kit AMS 700. Model: 72401850 - Accessory Kit AMS 700. Anvisa Registration n ° 80219980004.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por AMS American Medical Systems do Brasil Prod. Urol. Ginec. Ltda; AMS American Medical System Inc..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1533
  • Fecha
    2015-03-13
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    Between April 2014 and August 2014, AMS received 17 complaints for the 5325 devices sold by Keith Needles regarding bending or breaking during use. A review of the history of complaints (August 1, 2012 through July 31, 2013) showed that AMS received a complaint regarding the Keith Needle fold and the complaint regarding the break / component - tool or component / accessory kit break among the 13,708 devices sold. While a preliminary investigation was performed on the observed complaining trend, the supplier noted that the needles associated with 3 batches appeared to not have a fold close to the orifice. This fold is present in the other lots that have been historically supplied by the same supplier. The preliminary AMS investigation indicates that the missing fold may lead to a higher index of folded or broken needles. The supplier decided to recall the 3 batches of Needle Mounts with Fold 84413952.
  • Causa
    Absence of bending near the needle hole can lead to a higher index of folded or broken needles.
  • Acción
    The company is collecting the product from the market.