Alerta De Seguridad para ACCU CHEK PERFORMA - Registration with Anvisa nº 10287410617

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Roche Diagnóstica Brasil Ltda..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    971
  • Fecha
    2009-07-23
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    This situation is restricted to samples of neonates (born up to 28 days) when blood glucose tests are performed with this system and compared with other glucose monitors. In results below 50mg / dl (or 2.8mmol / L), results with negative deviation (lower values) were identified from -7 to - 12 mg / dl (-0.4 to -0.7mmol / L). Internal investigations are being conducted by the company in order to determine the reasons that lead to this deviation of performance. #### 15/10/2010 - After the investigations, the optimized product reached all acceptance criteria, in order to meet all necessary requirements for neonatal dosing. Starting with the 321001 test strip lot, all ACCU-CHEK Performa strips will include the improved calibration process, with performance characteristics for capillary, venous, arterial, and neonatal specimen.VIDE APPENDICES ####
  • Causa
    It was observed that when blood samples of neonates were collected by accu chek performa, low blood glucose values ​​were obtained. see clarifications as the termination of alert.
  • Acción
    The actions recommended to the users of the product are the following: Immediate interruption of the use of this product in neonates.