Alerta De Seguridad para Angiography Equipment Allura Xper Brand Philips. Models: Allura Xper FD10 Ceiling / Allura Xper FD10 Floor. Anvisa Registry n ° 10216710153.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Philips Medical Systems Ltda.; Philips Medical Systems Nederland B.V..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1699
  • Fecha
    2015-11-06
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    The severity was classified by Philips as "results in less than moderate damage or does not result in damage" and the probability of occurrence was classified as a "good chance of occurrence". The conclusion of the risk analysis performed by the company was acceptable. No injury was reported attributed to the problem so far according to the company. #### UPDATED ON 7/18/2017, the company sent a completion report of the field action proving the accomplishment of correction, as planned.
  • Causa
    Intermittent electronic defect in the product. under certain circumstances, a software error may lead to a situation where the five-minute fluoroscopy beep is not emitted as required in 21cfr1020.32 (h) (2) (ii) and iec 60601-2-54 , clause 203.6.2.1.C.
  • Acción
    Field correction (software update). Failure to emit the audible alarm occurs very intermittently. The user should always observe the real-time dose information and cumulative fluoroscopy time provided by the system. The fault condition is reset when the case of a new patient is started or when the system is reset.