Alerta De Seguridad para BARD TRUGUIDE COAXIAL NEEDLE FOR BIOPSY, model C2016B, registration 80689090075, hazard class II, lots: REWL0924, REXA1695, REXB1529, REXC1632.

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Bard Brasil Indústria e Comércio de Produtos Médicos Ltda.; Bard Peripheral Vascular, Inc..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1819
  • Fecha
    2016-02-12
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    The potential risk associated with attempting to proceed with coaxial placement, using the sharpened stylus rather than the optional blunt tip stylus, may be variable depending on the anatomical location of the biopsy and may cause an additional unintended risk to tissue, vessels and other organs. ### UPDATED ON 10/26/2017, the company presented the completion report of the field action proving the sending of the security warning to the client with evidence of science and all actions completed.
  • Causa
    The coaxial biopsy needle may inadvertently have a blunt-tipped stylet of incorrect size that will not pass through the coaxial. specifically the product should contain a blunt tip stylus of 19 gauge and actually contains a 17 gauge stylus.
  • Acción
    Gathering and destruction. Evaluate inventory and do not use or distribute any affected product. If the affected product has been safely used, no further action is required.