Alerta De Seguridad para CELL-DYN 1700, Registration 10055310574 - CELL-DYN 1700 CS, Registration 10055310576 - CELL-DYN 1600 CS (120v) Registration 10055310125 - CELL-DYN 1600 CS (220v) Registration - 10055310125 - CELL-DYN 1400, c / RS 232 Registration 10055310122 - CELL-DYN 1400 CS Registration 100553105122 and CELL-DYN 1400 Registration 10055310122

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Abbott Laboratórios do Brasil Ltda - Divisão Diagnósticos.

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    797
  • Fecha
    2004-08-18
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    Note the existence of abnormal platelet histogram with an MCV <7) fl parameter, as it may be a case of microcytosis and thrombocytopenia, with no URI alert message appearing. According to the results, of red blood cells and platelets, by an alternative manual or automatic method, according to the laboratory procedure. The platelet histogram demonstrates a normal or typical platelet histogram, which should be considered as the visual validation criterion of the result. As well as, patient boundaries indicate an abnormality of outcome, and are used to repeat and confirm by alternative methods manual or automated, according to the laboratory procedure. You are requested to: Keep a copy of the communication received in your files and attached to the CEL-DYN 1400, 1600 and 1700 instrument manuals. Forward a copy of the communication to the physician. Fill in the RESPONSE PROTOCOL and send the Customer Service Center. Any additional questions or clarifications, please contact the Customer Service Center, Telephone 0800 119099
  • Causa
    Software-related occurrence: the algorithm that governs the uri signaling mechanism inhibits the display of the signal for blood samples that are characterized by low values ​​in certain areas of sample histogram. a blood sample with combination of thrombocytopenia and microcyst of red blood cells is not likely to trigger uri signaling. the design of the software in combination with the characteristics of certain patient samples is the cause of the occurrence.
  • Acción
    Investigation into the cause of occurrence was completed. An initial preventive action was implemented. The initial preventive action consists of a review of the impact of this occurrence on other CELL-DYN Séire 1000 platforms. The CELL-DYN 1400 and CELL-DYN 1600 analyzers have essentially the same platelet algorithm as the CELL-DYN 1700. They may, however , exhibit the same URI signaling behavior for platelets with respect to thrombocytopenic / microcytosis samples, such as those of patients with thalassemia and idiopathic thrombocytopenic purpura.