Alerta De Seguridad para Commercial Name: Allura Xper Angiography Equipment; ANVISA registration numbers: 10216710153; Risk class: III; Aura affected: Allura Xper FD20; Serial numbers affected: 2352 (which is in Ribeirão Preto-SP).

Según Agência Nacional de Vigilância Sanitária (ANVISA), este evento ( alerta de seguridad ) involucró a un dispositivo médico en Brazil que fue producido por Philips Medical Systems Ltda.; Philips Medical Systems Nederland B.V..

¿Qué es esto?

Las alertas proporcionan información importante y recomendaciones sobre los productos. Aunque se haya emitido una alerta, esto no significa necesariamente que el producto se considera peligroso. Las alertas de seguridad, dirigidas a trabajadores de la salud y a usuarios, pueden incluir retiro de equipos. Pueden ser escritas por los fabricantes, pero también por funcionarios del área de salud.

Más información acerca de la data acá
  • Tipo de evento
    Safety alert
  • ID del evento
    1859
  • Fecha
    2016-04-01
  • País del evento
  • Fuente del evento
    ANVISA
  • URL de la fuente del evento
  • Notas / Alertas
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notas adicionales en la data
    Philips will contact you for scheduling and Philips maintenance engineers will install Software that solves the audible alarm problem. If you need any additional information or support regarding this issue, please contact our Solution Center at 0800-701-7789. Reception from Mon to Sat from 07 to 19h.
  • Causa
    Philips healthcare has identified, through customer complaints and internal testing, an intermittent electronic defect in the product. under certain circumstances, a software error may lead to a situation where the five-minute fluoroscopy beep is not emitted as required in 21cfr1020.32 (h) (2) (ii) and iec 60601-2-54 , clause 203.6.2.1.C. no injuries attributed to the problem were reported. failure to emit the audible alarm occurs very intermittently. the user should always observe the real-time dose information and cumulative fluoroscopy time provided by the system. the fault condition is reset when the case of a new patient is started or when the system is reset.
  • Acción
    Field Action Code: FCO72200309. Field Correction through Software Installation that solves the sound alarm problem (Software update R8.1.15 to R8.1.17.2) //////////// Recommendations to users and patients: Non-issue of the audible alarm occurs very intermittently. The user should always observe the real-time dose information and cumulative fluoroscopy time provided by the system. The fault condition is reset when the case of a new patient is started or when the system is reset.